Integrated IPX1 to IPX9K Water Ingress Test Chamber
This all-in-one IP waterproof test chamber is a versatile solution that combines IPX1/2 drip tests, IPX3/4/4K water spray tests, IPX5/6/6K strong water jetting tests, and IPX9K high-temperature, high-pressure water spray test in one unit. It accommodates different waterproofing standards for medical devices, consumer electronics, automotive parts, and other products, helping you reduce procurement costs and save valuable lab space.
The equipment is constructed with welded rectangular heavy-duty steel tubs and cold-rolled steel panels, and the chamber features a 304 stainless steel upper test chamber to house components for various test ratings. A square drip box with evenly spaced needles is mounted at the ceiling to simulate light rain for IPX1/IPX2 testing. In the center, a stainless steel turntable sits on a vertical rod anchored at the bottom of the chamber, holding your specimen. The turntable’s rotation speed and tilt angle are adjustable to fit your test needs.
In the middle area, a semi-circular oscillating tube with a defined number of nozzles, spaced at set intervals, enables IPX3/IPX4/IPX4K testing at 120 degrees or 360 degrees of oscillation. The tube ends attach to the left and right walls of the inner chamber. A swing arm in the inner corner carries four high-pressure IPX9/K nozzles at specified angles and spacings. These nozzles conform to relevant standards and can spray high-temperature water (80 ± 5 °C) at controlled pressures and intervals throughout the test. The IPX9 swing arm rotates horizontally and can be tucked against the wall when not in use.
Inside the innermost wall, a rectangular or circular opening connects to a 2.5 m water guide conduit. The two jet nozzles for IPX5, IPX6, and IPX6K are mounted on the opposite side of this conduit. The nozzles can be fixed or automatically swung up and down. During testing, a water jet is delivered through the guide conduit and directed at the rotating specimen for a defined period.
Customization options:
- Chamber size can be tailored to your specimen dimensions.
- Flexible configurations allow combinations of IPX ratings from IPX1 to IPX9 integrated into a single chamber per your test requirements.
- The sample can be powered on and off as needed during the test.

IPX1~4 combined test chamber

IPX3~6 combined test chamber

IPX1~6 combined test chamber

IPX3~9 combined test chamber
AmadeTech’s IP integrated rain test chamber comes with a colorful touchscreen and a built-in PLC. You manage all test controls from this single screen, which saves space and makes operation easier. You can tap the icons to access different test ratings and enter the required parameters, then monitor and control the test status from the same interface.
For each test item, the required flow rate is set with adjustable electronic flow meters. Alternatively, the integrated test chamber can be equipped with manually adjustable meters if there is sufficient space. Every waterproof test rating is independently controlled and won’t affect the others. When you start the machine, it follows the standard procedure and stops automatically when the test is complete.
All data and progress are displayed in real time on the screen, so you can easily view and monitor the test as it runs, ensuring a smooth, steady process.
IPX1-IPX4 combined test chamber
IPX1-IPX6 combined test chamber
IPX3-IPX9K combined test chamber
Test Standards
IEC 60529, ISO 20653, DIN 40050-9, IEC 60598-1, IEC 60068-2-18, IEC 60034-5, etc.
Technical Parameters
Model | AT-C1005 series | |||
General parameters | Main material | #304 Stainless steel | ||
Controller | 7” touch screen + PLC | |||
Interior chamber dimensions | W1300 x D1300 x H1300 mm (Optional) | |||
Water supply | Tap water (Recyclable) | |||
Sample power supply | available | |||
Turntable rotation | Forward / reverse alternately | |||
Turntable speed | 1 ~ 5 rpm adjustable | |||
Turntable height | Adjustable | |||
Machine power | 3-phase, AC 380V, 50Hz or specified by user | |||
IPX1/2 parameters | Drip area | 1000 x 1000 mm (optional) | ||
Needle diameter | 0.4 mm | |||
Needle spacing | 20 mm | |||
Water flow rate | 1 ~ 5 rpm adjustable | |||
Turntable angle | 0° or 15° | |||
Test duration | 10 min adjustable | |||
IPX3/4/4K parameters | Oscillating tube diameter | 600 mm (optional) | ||
Inner diameter of oscillating tube | 15 mm | |||
Open hole aperture | 0.4mm (IPX3 & IPX4), 0.8mm (IPX4K) | |||
Open hole spacing | 50 mm | |||
Swing angle of oscillating tube | IPX3 ± 60°(120°) IPX4, IPX4K ± 180° (355° in actuality) | |||
Swing speed of oscillating tube | 1 second/60° 4 seconds for 2 x 120° 12 seconds for 2 x 360° | |||
Water flow rate | 0.1L/min per open hole (IPX3/IPX4) 0.6L/min per open hole (IPX4K) Electronically adjusted | |||
Test duration | 10 min adjustable | |||
IPX5/6/6K parameters | Nozzle diameters | Dia. 6.3mm (IPX5,IPX6K) Dia. 12.5mm (IPX6) | ||
Water jetting distance | Appr. 2.7 meters | |||
Nozzle swing mode | Fixed or automatically swing up and down under control | |||
Water flow rate | IPX5: 12.5 L/min±5% IPX6: 100L/min±5% IPX6K: 75 L/min±5% Electronically adjusted | |||
Test duration | Min. 3 min adjustable | |||
IPX9/9K parameters | Test water temperature | 80℃±5℃ | ||
Number of nozzles | 4 pcs | |||
Spray angle | 0°, 30°, 60°, 90° | |||
Water Impact force | 0.9 ~ 1.2N (Calibration device included) | |||
Distance between nozzles and sample | 100 ~ 150mm adjustable | |||
Water flow rate | 14 ~ 16L/min (Electronically adjusted) | |||
Test duration | 30 seconds per position, 120 seconds in total for one cycle, or set arbitrarily | |||
Test Chamber Features
- Constructed with heavy-duty steel tubes and 304 stainless steel plates.
- IPX1 to IPX6K and IPX9/K capabilities can be integrated into a single test machine, saving lab space.
- You can flexibly select test items inside the chamber, and the chamber size can be customized to your needs.
- A 7-inch touch screen lets you set parameters, control tests, and view data all in one place.
- The machine automatically stops when the predetermined test item is completed.
- Each test item is controlled independently from the others.
- Water flow and pressure are individually adjustable for each test item.
- The turntable rotation speed is adjustable to fit different tests.
- A built-in water tank enables water recycling throughout testing.
- No waterproofing requirements at the test site are needed.
- If needed, you can energize the sample during the test.
- The needles and nozzles are easy to swap.
- The system includes protections such as low-water shutoff, short-circuit, and overload safeguards.
- It’s easy to use and maintain for operators.
FAQs
Request a Quote Now
Please feel free to contact us for more details on the product, price, lead time, payment terms, shipment methods, etc. AmadeTech sales engineers will respond within one working day.